20 Keysight E5052B Signal Source Analyzer
Installation
Power Supply
2-
Power Supply
Before turning on the E5052B power, check the following.
Verification of the Power Supply
Confirm that the power supplied to the E5052B meets the following
requirements:
Verification and Connection of Power Cable
The three-wire power cable attached to the E5052B has one wire serving as a
ground. Using this power cable allows the E5052B to be grounded, thereby
protecting you against electrical shock from the power outlet.
Step 1. Confirm that the power cable is not damaged.
Step 2. Use the supplied cable to connect between the power cable receptacle
(Figure 2-6 on page 24) on the rear panel of the E5052B and a three-wire
power outlet with the grounding prong firmly connected in the ground slot.
Figure 2-4 shows the power cable options.
Requirements
Voltage
90 to 132 VAC or 198 to 264 VAC
1
Frequency 47 to 63 Hz
Maximum power consumption 500 VA
1. Switched automatically by the E5052B in conformity to the voltage used.
NEVER use a power cable showing any sign of damage. Faulty cables can
cause electrical shock.
Use the supplied power cable with grounding wire to securely ground the
E5052B.