Keysight E5052B Signal Source Analyzer 45
Troubleshooting
System Recovery
HDD Revision: NM320 or below
Step 1. Shut down the E5052B.
Step 2. Disconnect all of the USB device from the USB ports.
Step 3. Connect the keyboard to the E5052B.
Step 4. Press the standby switch of the E5052B to turn it on.
Step 5. When the screen as shown in the figure below appears, press of the
keyboard.
You need a keyboard for this operation.
Press F11 only once. If you press F11 twice or more and a screen which is
not mentioned in this manual appears, reboot the instrument and restart
the process from the first.
After several seconds, the next screen appears automatically even if you
do not press any key, so do not miss it.
After Step 5, if the E5052B freezes with the cursor does not blink, turn the
power off and remove unnecessary USB devices. Then, replace the
keyboard with a PS/2 keyboard to execute the process again.
F11
e5071cse1055
Starting Acronis Loader...
Press F11 for Acronis Startup Recovery Manager...