Keysight X-Series Signal Generators Error Messages 21
Error Messages
List of Error Messages
-310 System error; Cannot change manual point until list mode error
condition cleared.
An error is keeping the sweep/list from being able to set the
frequency and/or power. Until the problem is addressed, the manual
point cannot be changed.
-310 System error; Communication failure with Application or CalTest
harness.
A failure occurred when attempting to communicate with the software
module responsible for digital modulation applications or
calibration/self test module. Cycle the instrument's power. If this
issue persists, please report it to the service office.
-310 System error; Could not create percent complete window semaphore.
When trying to initialize a percent complete Window the needed
semaphore could not be created. Report this error to service
office.
-310 System error; Could not limit frequency settings. Please preset.
Based on the current frequency settings (multiplier, offset, etc.)
the signal generator cannot successfully limit the frequency.
Preset the signal generator to insure the firmware will return to a
valid state.
-310 System error; Critical digital module error. Report this error to
service office.