Chapter 3 113
Instrument Messages
Condition Errors 201 to 299, Signal Integrity
207
(cont
Burst Not Found
(cont.)
E GSM: Data was acquired but a GSM burst was not
found, with the timeslot mode disabled.
NADC, PDC: A valid burst is not found when the
Device is MS.
1xEV-DO: Data was acquired but a 1xEV burst was not
found, with the timeslot mode disabled.
Bluetooth: The burst that has been found does not
correspond to the currently selected Bluetooth packet
type (the burst length may be too short).
WLAN: The instrument cannot find a valid WLAN
burst. You may need to extend the search length.
209 Timing Error E
209 Timing Error:No time
ref pilot burst
E The pilot burst used for time reference is not active.
211 Carrier(s) incorrect
or missing
E
213 Freq Out of Range E
213 Freq Out of Range;
External LO
E One or more external LO frequencies are out of range.
Check that the LO frequency limits are set correctly
and check the entered measurement frequencies and
measurement mode.
213 Freq Out of Range;
System input (IF)
E One or more system input frequencies are out of range.
If using a frequency list, check that all entries are valid
for current measurement mode.
215 Sync Error E W-CDMA: Cannot sync DPCCH pilot.
Cannot synchronize measurement with DPCCH pilot
for Power Control measurement, because the pilot
signal cannot be found. Make sure DPCCH is present
in the W-CDMA uplink signal, and that the slot format
and scramble code are set correctly.
215 Sync Error;Midamble
sync fail
E Failed to find the uplink slot, which caused the
synchronization with the midamble to fail.
215 Sync Error;No freq
ref pilot burst
E The pilot burst used for frequency reference is not
active.
215 Sync Error;No pilot
burst
E There is no Pilot burst detected.
215 Sync Error;Preamble
length zero
E Burst type is "Data" or "Preamble" and the
measurement cannot find a Preamble
215 Sync Error;Sync code
not found
E Synchronization code is not found in the measured time
slot.