Chapter 13 267
Assembly Replacement Procedures
Front Frame Assembly
Front Frame Assembly
Removal
NOTE Make sure any connectors on the front panel are removed.
1. Remove the instrument outer case. Refer to the Instrument Outer Case removal
procedure.
NOTE If the analyzer has Option PRC, Portable Configuration, and you want to remove
the bail handle and bumpers from the front frame, refer to the “Bail Handle
Removal (Option PRC)” section on page 269.
2. Refer to Figure 13-29. Using the T-10 driver, remove the eight screws (1),
four on each side, to detach the Front Frame Assembly from the chassis.
3. Refer to Figure 13-30. Pull the Front Frame Assembly carefully away from the
chassis. Remove the ribbon cable W1 from the mother board.
Figure 13-29 Front Frame Assembly Removal