Chapter 15 293
Functional Tests
Test Equipment
Signal Source
All Synthesized Sweeper Frequency: 10 MHz to 26.5 GHz
Harmonic level:< − 40 dBc
Amplitude range: 10 to − 20 dBm
Frequency Accuracy: 0.02%
PSG
a
83630B,
83640B,
83650B
Power Meter
All Power Meter Power Reference Accuracy: ±1.2%
Compatible with power sensor
E4418B E4419B
All Power Sensor Frequency Range: 50 MHz to
3.66 GHz
Amplitude Range: −70 to –10 dBm
8481D 8487D
8485D
P03 Low Power Sensor Frequency Range: 50 MHz to
3.6 GHz
Amplitude Range: –70 to –10 dBm
8481D 8485D
8487D
Pxx other
than P03
Low Power Sensor Frequency Range: 50 MHz to
26.5 GHz
Amplitude Range: –70 to –10 dBm
8485D 8487D
Oscilloscope
All Agilent Infiniium Oscilloscope Cutoff Frequency: 50 MHz
Rejection at 65 MHz: > 40 dB
Rejection at 75 MHz: > 60 dB
Rejection at 80 MHz: > 60 dB
Frequency: 10 MHz to 26.5 GHz
54800B
Miscellaneous Equipment
All Filter, 50 MHz Low Pass Cutoff Frequency: 50 MHz
Rejection at 65 MHz: > 40 dB
Rejection at 75 MHz: > 60 dB
Rejection at 80 MHz: > 60 dB
Frequency: 10 MHz to 26.5 GHz
0955-0306
All Power Splitter Nominal Insertion Loss: 6 dB
Tracking Between Ports: < 0.25 dB
11667A 11667B
All Termination, 50Ω Type N (m) Connector
Frequency: 30 Hz to 26.5 GHz
909A Option 012
a. PSG model numbers: E8244A, E8254A, E8247C Option 520, E8247C Option H31 or 540,
E8257D Option 520, E8257D Option 550, E8257D Option 567, E8267D Option 544.
Analyzer
Option
Item Critical Specifications
Recommended
Agilent Model
Alternate
Agilent Model