30 Chapter 1
Overview
Service Equipment You Will Need
Microwave Power Sensor
(2 required)
Frequency Range: 50 MHz to 26.5 GHz
Amplitude Range: -30 to +20 dB
Zero Set: ± 50 mW
Zero Drift: < ± 10 nW
Measurement Noise: < 110 nW
Cal Factor Uncertainty: < 2.3%
VSWR: 50 MHz to 100 MHz: 1.15:1
100 MHz to 2 GHz: 1.10:1
2 GHz to 12.4 GHz: 1.15:1
12.4 GHz to 18 GHz: 1.20:1
18 GHz to 26.5 GHz: 1.25:1
Input Connector: 3.5 mm (m)
8485A A, P
Low Power Microwave Power
Sensor
(2 required)
(for Option P08, P13,
or P26 only)
Frequency Range: 50 MHz to 26.5 GHz
Amplitude Range: −70 to +20 dBm
Zero Set: ± 50 pW
Zero Drift: < ± 4 pW
Measurement Noise: < 4 pW
Cal Factor Uncertainty: < 2.6%
VSWR: 50 MHz to 100 MHz: 1.15:1
100 MHz to 2 GHz: 1.10:1
2 GHz to 12.4 GHz: 1.15:1
12.4 GHz to 18 GHz: 1.20:1
18 GHz to 26.5 GHz: 1.25:1
Input Connector: 3.5 mm (m)
8485D A, P
RF Power Sensor
(2 required)
(CXA only)
Frequency Range:
9 kHz to 18 GHz
E9304A
(Option H19)
E9304A
(Option H18)
A
Power Sensor Cable
(2 required)
11730A A, P
Standards
Frequency Standard Frequency: 10 MHz
Accuracy: < ±1 e10
-10
Symmetricom
5071A
Agilent 5061B,
5071A
A, P
50 MHz,
−25 dBm Calibrator
Frequency Drift: < 2.5 kHz
Typical VSWR: 1.06:1
Output Power Variation: ±.004 dB
Total Harmonic Content: -45 dBc
Z5602A
Opt H51
A
Attenuators
10 dB
Step Attenuator
Range: 0 to 110 dB
Accuracy:
Characterized by standards lab to:
±0.005 + 0.005/10 dB step
d
Calibrated at 50 MHz
VSWR: at 50 MHz: ≤ 1.05:1
8496G 8496H P
1dB
Step Attenuator
Range: 0 to 11 dB
Accuracy:
Characterized by standards lab to: ±0.005 dB
d
Calibrated at 50 MHz
VSWR: at 50 MHz: ≤ 1.05:1
8494G 8494H P
Table 1-1 Required Test Equipment
Instrument Critical Specifications Recommended
Agilent Model
Number
Alternative
Agilent Model
Number
Use
a