Chapter 15 301
Functional Tests
Amplitude Accuracy at 50 MHz
Figure 15-4 Amplitude Accuracy Test Setup
Procedure
1. Zero and calibrate the power meter.
2. Configure equipment as shown in Figure 15-4, with the power splitter
connected directly to the signal analyzer input through the adapter.
CAUTION To minimize stress on the test equipment connections, support the power sensor.
3. If the auto alignment for the analyzer has not been performed within the past
24 hours, press
System, Alignments, Align Now, All to perform the auto
alignment routine.
4. Press
Mode, Spectrum Analyzer, Mode Preset on the analyzer.
5. Set up the synthesized sweeper by pressing:
CW, 50, MHz
Power Level
, -4, dBm
RF
(On)
6. Set up the signal analyzer by pressing:
Input/Output, More, Freq Ref In, External
FREQ Channel, Center Freq, 50, MHz
SPAN X Scale
, 2, MHz
AMPTD Y Scale
, Attenuation, Mech Atten, 10, dB
Input/Output
, RF Input, RF Coupling, select DC
Sweep
/Control, Sweep Setup, Swp Time Rules, SA - Accuracy
Meas Setup
, Average/Hold Number, 20, Enter
Trace/Detector
, Trace Average
Single