82 Chapter 3
Instrument Messages
-300 to -399, Device-Specific Errors
-310 System error; The
configured PTP hardware
driver could not be
instantiated.
The PTP driver failed on initialization.
-310 System error; The PTP
hardware driver reported
a configuration error.
Failure in the execution of the PTP driver. The most likely
cause of this error is a mismatch between versions of the PTP
driver and the LXI middleware.
-310 System error; The PTP
ordinary clock reported
a configuration error.
Failure in execution of the DMC libraries’ PTP controller.
-310 System error; The
Trigger alarm delayed
LAN event was not
scheduled due to an
existing pending event.
Delayed LAN events cannot occur too close together (within
20 ms).
-310 System error; The
Trigger alarm delayed
LAN event was not
scheduled due to a
conflict with an
existing scheduled
alarm.
Delayed LAN events cannot occur too close to a scheduled
Alarm (within 20 ms).
-310 System error; The
Trigger alarm was not
scheduled due to a
conflict with an
existing scheduled
alarm.
Alarms cannot be scheduled to happen too close together
(within 20 ms).
-311 Memory error There is a physical problem with the instrument memory, such
as a parity error.
-312 PUD memory lost Protected user data saved by the *PUD command has been
lost.
-313 Calibration memory lost The nonvolatile calibration data used by the *CAL? command
has been lost.
-314 Save/recall memory lost The nonvolatile data saved by the *SAV? command has been
lost.
-315 Configuration memory
lost
The nonvolatile configuration data saved by the instrument has
been lost.
-320 Storage fault; A problem was found while using data storage. The error is not
an indication of physical damage or failure of any mass storage
element.
-321 Out of memory An internal operation needed more memory than was
available. Report this error to the nearest Agilent Technologies
sales or service office.