EasyManua.ls Logo

Keysight N9000A - Page 83

Keysight N9000A
322 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 83
Instrument Messages
-300 to -399, Device-Specific Errors
-330 Self-test failed A self-test failure occurred. Report this error to the nearest
Agilent Technologies sales or service office.
-340 Calibration failed The instrument requires an Align All Now. Restore the
alignment by pressing System, Alignments, Align All Now.
-340 Calibration failed; no
signal detected; <port>
The calibration for the <I | I-bar | Q | Q-bar> port did not
succeed because the calibration signal was not detected. Check
that the < I | I-bar | Q | Q-bar > port is properly connected to the
Cal Out port.
-350 Queue overflow An error occurred that did not get put in the error queue
because the queue was full.
-360 Communication error There was a problem with instrument remote communications.
The exact problem cannot be specifically identified.
-360 Communication error; SNS
data read failure.
Disconnect then
reconnect SNS
The Agilent Smart Noise Source connected to the CXA has
failed to be read by the application. Please disconnect and
reconnect the SNS. If this continues to fail, then the SNS may
have had its EEPROM corrupted or another hardware fault
exists. Check SNS on another instrument, NFA and ESA are
also SNS compatible instruments. Check the device is not an
Agilent power sensor which uses the same cable interface.
-361 Parity error in program
message
A parity bit was not correct when the data was received. For
example, on a parallel port.
-362 Framing error in program
message
A stop bit was not detected when data was received. For
example, on a remote bus port.
-363 Input buffer overrun A software or hardware input buffer on a port overflowed with
data because of improper or nonexistent pacing.
-365 Time out error There was a time-out problem in the instrument. The exact
problem cannot be specifically identified.