86 Chapter 3
Instrument Messages
-221 Settings Conflict Errors
-221 Settings conflict; Corrections
with different antenna units
not allowed
When a correction with antenna units is turned on, that
is the only unit allowed. You can have two sets of
antenna corrections turned on, but only if they have the
same units.
-221 Settings conflict; Desired
preamp setting is not
available
-221 Settings conflict; Destination
trace for Trace Math cannot be
a trace operand
The resulting trace data (from doing a trace math
function) cannot be put into the any of the traces that are
being used by the math operation.
-221 Settings conflict; Downconv
only available when DUT is
Amplifier
SCPI only message. The System Downconverter can
only be set to ‘On’ when the DUT type is amplifier.
Change DUT type to Amplifier if the System
Downconverter is required.
-221 Settings conflict; EDGE EVM
only supports EDGE TCH burst
type.
-221 Settings conflict; Electronic
attenuator is disabled
You are using the mechanical attenuator, and have not
enabled the electronic attenuator. You cannot set the
value of the electronic attenuator because it
automatically sets/changes when enabled.
-221 Settings conflict; Electronic
attenuator is not available
above 3.6 GHz
The maximum frequency of the electronic attenuator is
3.6 GHz. This is because of switching capacitance.
-221 Settings conflict; Electronic
attenuator unavailable in
current state
-221 Settings conflict; Electronic
attenuator unavailable with
Preamp on
The internal preamp is on. Electronic attenuator cannot
be used while you are using the internal preamp.
-221 Settings Conflict; FAST method
can only be used while Radio
Std is W-CDMA
-221 Settings conflict; Feature not
supported for this
measurement.
Some functionality is available in one measurement, but
not in another. (See the measurements under the Meas
key.) This error occurs if you send a SCPI command or
push a gray-out key that is not available in the current
selected measurement.
-221 Settings conflict; FFT IF Gain
High not available when Swept
IF Gain = Manual Low
When Swept IF Gain is manually set to Low, you cannot
set the FFT IF Gain to High because that would make
the Reference Level couplings wrong in FFT mode.
-221 Settings conflict; FFT method
is unavailable for level
gating
If you are using level gating, you cannot select the FFT
Gate Method.