Chapter 3 87
Instrument Messages
-221 Settings Conflict Errors
-221 Settings conflict; FFT sweep
type is not available while in
Gated LO
The gated LO function turns the LO on and off as it
sweeps. So the FFT sweep type is not available if you
have selected gated LO.
-221 Settings conflict; FFT sweep
type is not available while in
Gated Video
The FFT sweep type is not available if you have
selected the gated video function.
-221 Settings conflict; Fixed
marker adjust not available
while Marker Function is on
If a Marker Function is on for a Fixed marker, the
marker's reported value is derived from the function.
Therefore, you cannot directly set the X or Y value of a
Fixed marker that has a marker function turned on.
-221 Settings conflict; Fixed
Marker Y value is not
adjustable with Normalize On
If Normalize is on the Amplitude scale is in dB units, so
adjusting the Y value of a Fixed marker is not possible.
-221 Settings conflict; Freq > 3.6
GHz unavailable while
electronic attenuator enabled
The electronic attenuator does not function above 3.6
GHz. So if you have that attenuator enabled, you cannot
change the center frequency so that frequencies above
3.6 GHz are displayed/measured.
-221 Settings conflict; Frequency
Offset is not available with
Segmented Sweep
The frequency offset feature cannot be used when you
are using the segmented sweep capability.
-221 Settings conflict; Frequency
Offset not available when
Frequency Scale is Log
The frequency offset feature cannot be used when you
have selected a log scale for the frequency axis.
-221 Settings conflict; Function
unavailable with MW Presel off
You cannot center or adjust the preselector because the
Microwave Preselector is currently off
-221 Settings conflict; Gate
control is Edge for Gated FFT
You cannot use level triggering to control the gate if you
are using the FFT gating method.
-221 Settings conflict; Gate
control must be Edge for this
Gate Source
You cannot use level triggering to control the gate when
you are using the currently selected gate source.
-221 Settings conflict; Gate is not
available when Marker Count on
The gate function cannot be used while you have
marker count turned on.
-221 Settings conflict; Gate Length
is not settable in FFT sweeps
The sweep time for FFT sweeps is set by the
calculations. So sweep time settings cannot be adjusted.
-221 Settings conflict; Gate Length
is not settable in FFT sweeps
The sweep time for FFT sweeps is set by the
calculations. So sweep time settings cannot be adjusted.
-221 Settings conflict; Gate Method
is not compatible with current
Sweep Type setting
If the Gate is On and you have the FFT Sweep Type
manually selected, then the Gate Method cannot be
selected.
-221 Settings conflict; Gated FFT
is not available while Sweep
Type is set to Swept
The gated FFT function is not available if you have
selected the swept type of sweep. You must be in the
FFT sweep type.