88 Chapter 3
Instrument Messages
-221 Settings Conflict Errors
-221 Settings conflict; Gated LO is
not available while Sweep Type
is set to FFT
The FFT sweep type moves the LO frequency in steps.
So the gated LO function is not available if you have
selected FFT sweep.
-221 Settings conflict; Gated Video
is not available while Sweep
Type is set to FFT
The gated video function is not available if you have
selected the FFT sweep type.
-221 Settings conflict; Incorrect
RBW for demod. Change RBW
-221 Settings conflict; Ind I/Q is
not available for this
measurement
The Independent I and Q setting is not available for the
current measurement. Only some measurements
(initially, only VXA) support this setting.
-221 Settings conflict; Invalid
trace number
The subopcode used to specify the trace number is
invalid for this measurement or query
-221 Settings conflict; Knob is not
available to modify this
function
You should select a specific value for this function. So
scrolling through values with the knob is not allowed.
-221 Settings conflict; Log Scale
Type is not available when
Detector Mode is Normal
-221 Settings conflict; Log Scale
Type is not available with
Demod View
The logarithmic x-axis scales are not available when
you have the demod view turned on.
-221 Settings conflict; Log Scale
Type is only available in
swept measurement
Logarithmic scaling can be used when making a swept
SA measurement. It is not available in the SA
measurement when you are using FFT sweeps.
-221 Settings conflict; Log Scale
Type not available when
Segmented Sweep is on
The logarithmic scales are not available while you are
using the segmented sweep capability.
-221 Settings conflict; Log Scale
Type only available when
Frequency Offset = 0 Hz
If you have any amount of frequency offset, then you
cannot use logarithmic scaling for the X axis.
-221 Settings conflict; Marker 1
Trace Update=off turns off
Signal Track
Signal Track not available unless the trace containing
Marker 1 is updating
-221 Settings conflict; Marker
cannot be relative to itself
A marker must be set relative to another marker, not to
itself.
-221 Settings conflict; Marker
Count is not available when
Gate on
The marker count function cannot be used while you
have gating turned on.