Chapter 3 89
Instrument Messages
-221 Settings Conflict Errors
-221 Settings conflict; Marker
Function is not available for
a Fixed marker
If a Marker is a Fixed type marker, the marker's value
does not change from when it first became fixed. You
cannot turn on or change a Marker Function because
there is no ongoing measurement data to use for the
marker function calculation.
-221 Settings conflict; Marker type
must be delta
MkrΔ->Span and MkrΔ->CF require that the selected
marker be a delta marker.
-221 Settings conflict; Marker->
command is not available with
segmented sweep
Most of the “Marker To” commands are not available
with segmented sweep turned on. So you cannot select
these functions.
-221 Settings conflict; Marker->
function is not available in
zero span
Most of the “Marker To” functions are not available if
you are in zero span (span = 0 Hz, or time domain). So
you cannot send the commands for these functions.
-221 Settings conflict; Meas Type
was changed to Examine for Exp
Avg Mode.
Average Mode has been changed to Exponential. Full
Meas Type is not available for Exponential Average
Mode therefore Meas Type has been changed to
Examine.
-221 Settings conflict; Meas Type
was changed to Full for Repeat
Avg Mode
Average Mode has been changed to Repeat. Examine
Meas Type is not available for Repeat Average Mode
therefore Meas Type has been changed to Full.
-221 Settings Conflict; MinPts/RBW
limit not met
-221 Settings conflict; Mixer Type
is not available when in this
band
The Mixer Type selection is not available in the K, E,
W, F, D, G, Y, J band because those bands do not allow
Preselected Mixing.
-221 Settings conflict; Mkr -> CF
is not available when the
x-axis is time domain
The marker to center frequency functionality does not
work when the x-axis is in the time domain.
-221 Settings conflict; MS gain is
not available in this Mode
Mobile Station gain correction is not available in some
Modes, or in some measurements (for example, the SA
measurement).
-221 Settings conflict; No meas
frequencies are above 3.6 GHz
Optimize Preselector can only be performed on
frequencies in high band i.e. freqs above 3.6 GHz. The
current setup does not have input freqs (IF) in this range
so an Optimize Preselector cannot be performed.
-221 Settings conflict; Normal
detector is not allowed with X
scale is Log
The normal detector cannot be used when the x-axis
scale is logarithmic.
Why not? Are the results weird? Frequency slewed?
-221 Settings conflict; Normalize
is not available when Scale
Type = Lin
Normalize does not support Linear amplitude scale,
since the results are always presented as a dB ratio.