90 Chapter 3
Instrument Messages
-221 Settings Conflict Errors
-221 Settings conflict; Normalize
is not available while Demod
View is on
The normalization (correction) function cannot be used
if you are using the Demod View.
-221 Settings conflict; Normalize
is not available while Trace
Math is on
The Normalize function works by doing trace
manipulation. So if trace math is on you cannot turn on
normalization.
-221 Settings conflict; Option not
available
You have attempted to perform an action for which a
required option is not installed
-221 Settings conflict; Param only
available when DUT is a freq
converter
SCPI only message. The sideband and freq context
parameters are only available when a freq conversion
setup is in use. Change setup to contain a freq
conversion to use these parameters.
-221 Settings conflict; Param only
available when External LO
Mode is Swept
SCPI only message. This parameter is only available
when the LO mode is set to Swept. Change the LO
Mode to Swept.
-221 Settings conflict; Param only
available when External LO
Mode is Fixed
SCPI only message. This parameter is only available
when the LO mode is set to Fixed. Change the Freq
Mode to Fixed.
-221 Settings conflict; Param only
available when Frequency Mode
is Fixed
SCPI only message. This parameter is only available
when the Freq mode is set to Fixed. Change the Freq
Mode to Fixed.
-221 Settings conflict; Param only
available when Frequency Mode
is Swept
SCPI only message. This parameter is only available
when the Freq mode is set to Swept. Change the Freq
Mode to Swept
-221 Settings conflict; Param only
available when valid cal data
exists
SCPI only message. The ‘Apply Calibration’ parameter
is only available when the stored cal data matches the
current setup. Perform a fresh ‘Calibrate Now’ or
change setup such that current cal data is valid.
-221 Settings conflict; Parameter
currently is disabled
This parameter is grayed out (unavailable) in the current
context. Check the individual parameter
help/documentation for more information.
-221 Settings conflict; Preamp gain
is not available in this Mode
Preamp gain correction is not available in some Modes
or Measurements
-221 Settings conflict; Preamp
unavailable with electronic
attenuator on
The electronic attenuator is on. Internal preamp cannot
be used while you are using the electronic attenuator.
-221 Settings conflict; Reference
marker must be in same window
A delta marker and its reference must be in the same
window. This error occurs when you try to turn on a
delta marker who’s reference is in a different window.
-221 Settings conflict; Res BW
cannot be auto-coupled while
in Zero Span
The resolution bandwidth cannot be set to auto while
you are in zero span (time domain).