Chapter 3 91
Instrument Messages
-221 Settings Conflict Errors
-221 Settings conflict; Scale Type
= Lin is not available when
Normalize is on
Only the Log amplitude scale is available in Normalize,
since the results are always presented as a dB ratio.
-221 Settings conflict; Settings
conflict; Mask unavailable for
current Span. Increase to
display mask.
The current span setting is either narrower than the
mask width or so wide that there are too few display
points to allow the mask to be drawn. Increase or
decrease the span to display the mask.
-221 Settings conflict; Settings
conflict; Pre-trigger is
insufficient for demod.
Decrease Trig Delay.
-221 Settings conflict; Signal
Track is not available when
Freq Scale=Log
The signal tracking feature cannot be used when you
have selected a log scale for the frequency axis.
-221 Settings conflict; Signal
Track is not available with
Continuous Peak
The signal tracking feature cannot be used while you are
also using the continuous peak function.
-221 Settings conflict; Signal
Track is not available with
Segmented Sweep
The signal tracking feature cannot be used when you are
using the segmented sweep capability.
-221 Settings conflict; Signal
Track is only available in
Swept SA measurement
The signal track functionality can be used when making
a swept SA measurement. It is not available in the SA
measurement when you are using FFT sweeps.
-221 Settings conflict; Signal
Track is turned off when Zero
Span is selected
Signal Track is not available when you have selected
Zero Span. So if Zero Span is entered while in Signal
Track is On, Signal Track is turned off.
-221 Settings conflict; Span is not
available when Segmented Sweep
is on
Segmented Sweep uses multiple sweeps, each with its
own span setting. You cannot set the span when
segmented sweep is selected because the function does
not know which span to change.
-221 Settings Conflict; Span
limited to XXX
-221 Settings conflict; Span Zoom
is not available in Zero Span
Span Zoom does not work with a time domain x-axis.
You must select a span greater then 0 Hz.
-221 Settings conflict; Span Zoom
not available when Frequency
Scale Type = Log
The Span Zoom feature cannot be used when the X-axis
scale is logarithmic frequency.
-221 Settings conflict; Span Zoom
not available when Segmented
Sweep is on
Segmented Sweep uses multiple sweeps so it does not
know what sweep to zoom in on. Also, it cannot zoom a
span over a sweep boundary.
-221 Settings conflict; Step keys
are not available to modify
this function
You should select a specific value for this function. So
using the Up/Down step keys to scroll through values is
not allowed.