96 Chapter 3
Instrument Messages
-200 to -299, Execution Errors
-220 Parameter error A problem was found with a program data element. The exact
problem cannot be specifically identified.
-221 Settings conflict; There are many types of settings conflict errors. See section 3.5
for information about these errors.
-222 Data out of range; A data element was found but the instrument could not be set to
that value because it was outside the range defined for the
command. A descriptive message may be appended, such as
“clipped to upper limit”
-222 Data out of range; Invalid
list data
You tried to use a trace that has a number of sweep points that is
different from the current setting of sweep points.
-222 Data out of range; Two
entries already exist at
this x-axis value.
When entering values for limit lines, you cannot have more than
two y-axis (amplitude) values entered for a specific x-axis
(frequency) value.
-223 Too much data A data element (of block, expression, array type, or string type)
had more data then allowed by the command, or by the available
memory.
-223 Too much data; 200 spurs
found. Additional spurs
ignored.
There are too many spurs for the table (the limit is 200), and any
additional spurs that are found will be ignored.
-224 Illegal parameter value An exact data value (from a list of the allowed values) was
required - but not found. See the feature description for
information about the expected parameter values.
-224 Illegal parameter value;
<Value> invalid.
Fractional values are not
allowed.
The seconds parameter of an LXI time may not contain a
fractional portion. For example 123456789.0 is valid while
123456789.1 is not.
-224 Illegal parameter value;
<value> out of range.
The value does not fall in the valid range
-224 Illegal parameter value;
Exceeding the max list
length
The list parameters have a maximum allowed length. You are
trying to set a length longer than the maximum.
-224 Illegal parameter value;
Gated FFT is not available
while Sweep Type is set to
Swept
The gated FFT function is not available if you have selected the
swept type of sweep. You must be in the FFT sweep type.
-224 Illegal parameter value;
Gated LO is not available
while Sweep Type is set to
FFT
The FFT sweep type moves the LO frequency in steps. So the
gated LO function is not available if you have selected FFT
sweep.
-224 Illegal parameter value;
Gated Video is not
available while Sweep Type
is set to FFT
The gated video function is not available if you have selected the
FFT sweep type.