Chapter 3 99
Instrument Messages
-200 to -299, Execution Errors
-250 Mass storage error;
Different Antenna Unit
already in use
Attempt to import Corrections file with Antenna Unit that differs
from an in-use correction.
-250 Mass storage error;
Directory not found
The system cannot find the path specified.
-250 Mass storage error; Failed
to Load trace. Bad file
format.
The load trace operation could not be completed, as the input file
was not in the expected format. You can only load traces that
were previously saved using the 'Save Trace' feature.
-250 Mass storage error; File
<filename> wrong type
Attempt to import a data file that is not the proper type for this
operation.
-250 Mass storage error; File
<filename> and instrument
version mismatch
While opening a file, there was a mismatch between file version
or model number with instrument version or model number. The
import still tried to load as much as possible, but you should
check it closely.
-250 Mass storage error; File
contains incorrect data
for this operation
There is a mismatch between the file data type of the file
specified and the destination indicated. For example, a correction
set cannot be loaded/imported into a limit line.
-250 Mass storage error; File
empty
Cannot save trace because it contains no data. Check that the
trace is turned on and contains some valid data.
-250 Mass storage error;
Invalid register number
for *SAV or *RCL
You have used the *SAV command to save a state to a non
existent state register.
Or
You have used the *RCL command to recall a state register that
wasn’t previously saved with the *SAV command.
-250 Mass storage error; Lock
violation
The process cannot access the file because another process has
locked a portion of the file.
-250 Mass storage error; No
file names available
Attempt to use the auto file name generation when all 10,000 file
names are taken.
-250 Mass storage error; Open
failed
The system cannot open the device or file specified. This could
be because the storage media is full, or possibly due to a
filename error. If using an external storage devise, check that the
devise is properly formatted.
-250 Mass storage error; Read
fault
The system cannot read from the specified device.
-250 Mass storage error;
Register <number> empty
Attempt to recall a register with nothing in it
-250 Mass storage error;
Sharing violation
The process cannot access the file because it is being used by
another process.
-250 Mass storage error; Too
many open files
The system cannot open the file.