RF Section Troubleshooting
Troubleshooting
Input Attenuator B Power Level Verification
Press AMPTD, Attenuation, 0 dB. Remove output cable W9 from A10 (2).
Refer to Figure 4-13. Measure the 50 MHz calibrator signal on the output of
the attenuator using a functioning Spectrum Analyzer.
The level should be 25 dBm ± 2 dB as shown in Figure 4-15.
Figure 4-15 50 MHz Calibrator Signal on Output of Attenuator B