RF Section Troubleshooting
Troubleshooting
Press Mech Atten and enter 6 dB. The 50 MHz calibrator signal measured on
the functioning Spectrum Analyzer should measure 6 dB lower than the
previous step (~31 dBm) as shown in Figure 4-16.
Figure 4-16 50 MHz Calibrator Signal on Output of Attenuator B
(with 6 dB Attenuation)
Press Mech Atten and enter 10 dB. The 50 MHz calibrator signal measured on
the functioning Spectrum Analyzer should measure 10 dB lower than the first
step (~35 dBm). Continue testing all the attenuation settings as described in
Table 4-11.
Table 4-11
Attenuator Setting (dB) Expected Level
(dBm @ 50 MHz)
Tolerance
16 41 ±3 dB
20 45 ±3 dB
30 55 ±3 dB
40 65 ±4 dB
50 75 ±4 dB
60 85 ±4 dB
66 91 ±4 dB