Command Reference
CALCulate Subsystem
64
CALCulate Subsystem
This subsystem is used to perform post-acquisition data
processing. In effect, the collection of new data triggers the
CALCulate subsystem. In this instrument, the primary functions in
this subsystem are markers and limits. CALCulate subsystem
commands used for measurements in the MEAS menus are located
in "SENSe Subsystem“ on page 122.
Autotune
:CALCulate:AUTOtune
Enable the analzer’s auto tune function.
Key access:
NdBpoints
:CALCulate:BWIDth|BANDwidth:NDB <rel_ampl>
:CALCulate:BWIDth|BANDwidth:NDB?
Selects the power level, below the peak of the signal, at which the
signal bandwidth will be measured by the markers.
:CALCulate:BWIDth|BANDwidth[:STATe] must be ON.
*RST: –3 dB
Range: –80 dB to –1 dB
Remarks: Refer to :CALCulate:BWIDth|BANDwidth[:STATe] for an
explanation of this marker function.
Key access: > More > N dB Points