Installation and Service Guide N5251-90001 6-9
PNA Series Microwave Network Analyzer System
N5251A
Performance Tests and Checks
System Performance Verification
Figure 6-4 System Verification Configure Tab
9. Select the Run tab to return to the previous window.
The remaining i
nstructions assume Port 1 and Port 2 are being
verified. Modify the instructions appropriately for Port 3 and Port 4.
10. Click Run.
11. Enter the serial number for the 85059A kit. The other numbers are optional.
12. Click Continue.
13. As prompted, install the 1.0 mm female to female cable (8.8 cm, Keysight part number
11500-60001) onto the right test head (Port 2). See Figure 6-5 on page 6-10. This cable is
part of the 85059A 1.0 mm Calibration and Verification kit and is considered to be a test
port cable.