14 Error Messages
132 Keysight U2722A/U2723A Programmer’s Reference
Error Command errors (these errors set the Standard Event Status register bit #5)
–110 Command header error
An error was detected in the header.
–111 Header separator error
A character that was not a valid header separator was found in the command string.
–112 Program mnemonic too long
The header contains more than 12 characters.
–113 Undefined header
A command was received that is not valid. You may have misspelled the command or it may not be a valid
command. If you are using the short form of the command, remember that it may contain up to four letters.
–114 Header suffix out of range
The value of the numeric suffix is not valid.
–120 Numeric data error
Generic numeric data error.
–121 Invalid character in number
An invalid character was found in the number specified for a parameter value.
Example: *ESE #2
–123 Exponent too large
A numeric parameter was found where its exponent was larger than 32,000.
–124 Too many digits
The mantissa of a numeric parameter contained more than 255 digits, excluding leading zeros.
–128 Numeric data not allowed
A numeric parameter was received but a character string was expected.
–130 Suffix error
Generic suffix error.
–131 Invalid suffix
A suffix was incorrectly specified for a numeric parameter.
–134 Suffix too long
The suffix contains more than 12 characters.
–138 Suffix not allowed
A suffix is not supported for this command.
Table 14-2 Command errors (continued)