8 SENSe Subsystem
48 Keysight U2722A/U2723A Programmer’s Reference
SENSe:VOLTage[:DC]:APERture?
Syntax
SENSe:VOLTage[:DC]:APERture? (@<ch>)
This will query the sampling time for a single voltage measurement point. The
parameter has a unit of seconds.
Parameter
Remark
The aperture value is derived from the following equation.
Returned Query Format
<NR3>
The reading is in the form +1.000000E+01.
Example
SENS:VOLT:APER? (@2)
Typical Response: +2.000000E-02
See Also
SENSe:CURRent[:DC]:APERture?
SENSe:CURRent[:DC]:NPLCycles
SENSe:VOLTage[:DC]:NPLCycles
SYSTem:LFRequency
Item Type Range of values Defaul t value
<ch> NR1 1 through 3 Required parameter
Aperture
=
NPLCycles
Line frequency