Performance Verification and Adjustment 3
U2722A/U2723A Service Guide 29
7 When the voltage adjustment has completed, connect the
respective U2722A/U2723A output to the current terminal
of the 3458A. After the connection is established, click OK
to continue with current adjustment for channel 1.
8 After the voltage and current adjustments for channel 1
have completed, the 3458A will proceed with voltage and
current adjustments for channel 2 followed by voltage and
current adjustments for channel 3. Follow the instructions
on the message box to complete the adjustment operation.
9 When the adjustment operation has completed, a message
box as shown in the following will appear. Click OK to end
the adjustment operation.
10 A message box will appear as shown below prompting you
to power cycle the U2722A/U2723A for the adjustment to
take effect. To power cycle the U2722A/U2723A, power off
the U2722A/U2723A and restart it. Ensure that the
U2722A/U2723A is powered- off properly before restarting.