Using Device Clear to Halt Measurements
Device clear is an IEEE-488 low-level bus message which can be used to halt
measurements in progress. The status registers, the error queue, and all
configuration states are left unchanged when a device clear message is received.
Device clear performs the following actions:
All measurements in progress are aborted.
The U8480 Series returns to the trigger “idle state”.
The U8480 Series input and output buffers are cleared.
The U8480 Series is prepared to accept a new command string.