Performance Verification for Digital Generator and Digital Analyzer 4
Keysight U8903B Service Guide 69
Channel 2 vertical axis unit: Volt
Input channel 1 and 2: Enable
Function 1 to subtract channel 1 and channel 2
Function 1: Enable
Function 1 offset: 0
3 Configure the U8903B digital output channel to be tested as follows:
Digital output type: Balanced
Digital output voltage: 2.5 V
Digital output sampling rate: 48 kHz
Digital output AES audio resolution: 24 bits
AES output: ON
Digital output function: SINE waveform
SINE waveform frequency: 1 kHz
SINE waveform amplitude: 1 FFS
Digital output state: ON
4 Set the digital output sampling rate to F
SR
as given inTest record for balanced output
level accuracy verification” on page 141.
5 Set the digital output level to V
DUT
as given in “Test record for balanced output level
accuracy verification” on page 141.
6 Configure the DSO8064A oscilloscope as follows:
Time base scale: 1 / (Sampling rate × 32 × 2)
Channel 1 and channel 2 vertical scale value: refer to Table 4-4
Function 1 vertical scale range value: refer to Table 4-4
Function 1 offset: 0
Table 4-4 Channel 1 and channel 2 vertical scale and vertical scale range
V
DUT
(V)
CH1 and CH2 voltage per division (V) Function 1 vertical scale range (V)
0.3 0.025 0.4
0.5 0.0425 0.68
0.8 0.075 1.2
1 0.085 1.36
1.2 0.1 1.6
1.3 0.125 2
1.5 0.125 2