5-94 2G0947 - 31JAN08
Diagnostics
Condition Code: 28-501
Description
This error occurs when the temperature calculated by the PROCESSOR
MICROPROCESSOR is nearly zero. The temperature is calculated from the
voltage measured by the RTD. The most likely cause is a short in the RTD
circuit. The error clears only when power is cycled.
System Impact
The printer is unable to print films. The spooler is still able to accept and queue
jobs but new films will not be initiated. Films downstream of the affected
subsystem will be completed. All other films will be reprinted.
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service.
FE Action
Condition Code: 28-509
Description
This error occurs if the PROCESSOR RTD indicates a temperature lower than
the setpoint temperature when the allotted warm-up time has elapsed. The
allotted time is based on the initial PROCESSOR temperature. The error clears
only when power is cycled.
System Impact
The printer is unable to print films. The spooler is still able to accept and queue
jobs but new films will not be initiated. Films downstream of the affected
subsystem will be completed. All other films will be reprinted.
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service.
FE Action
Web Portal Message Log Message DICOM Message Severity
Processor: Internal Hardware
Failure
Processor calculates
temperature near zero. Probably
short in RTD circuit.
PROC DOWN 0-Fatal
Web Portal Message Log Message DICOM Message Severity
Processor Warm-up Failure Processor failed to warm up in
the specified time
PROC DOWN 0-Fatal