2G0947 - 31JAN08 5-95
Diagnostics
Condition Code: 28-510
Description
This error occurs when the temperature calculated by the PROCESSOR
MICROPROCESSOR is near its maximum value. The temperature is calculated
from the voltage measured by the RTD. The most likely cause is an open in the
RTD circuit. The error clears only when power is cycled.
System Impact
The printer is unable to print films. The spooler is still able to accept and queue
jobs but new films will not be initiated. Films downstream of the affected
subsystem will be completed. All other films will be reprinted.
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service.
FE Action
Web Portal Message Log Message DICOM Message Severity
Processor: Internal Hardware
Failure
Processor calculates
temperature reflecting the
maximum sensor reading.
Probably open in RTD circuit.
PROC DOWN 0-Fatal