Hardware Error Messages
AT000115-02 11
Section 3: Hardware Error Messages
“Z0 Sensor at Wrong State”
Scenario
Screen Carrier is stuck while the screen is being loaded. As a result the Screen Carrier doesn’t reach the Z0 Sensor
(J505) so that it doesn’t change its position from “1” to “0”.
Hardware and Software Response
[1] Screen unload is halted.
[2] Cassette is still locked in the Reader.
[3] CR Reader operating status changes to “State stalled”.
Fix
User Response
[1] Click [OK]; the CR Reader is halted. Status remains “state stalled”.
[2] Manually release the cassette using the Release knob under the cassette tray, and remove the cassette.
[3] Reset the CR Reader by turning it OFF and ON again.
[4] Insert the cassette again; does the error repeat?
Service Response
[1] Remove the CR Reader cover.
[2] Is the Screen Carrier stuck?
Yes No
Contact Service Continue using the CR Reader
Yes No
a. Manually move the Screen Carrier to its homing position.
b. Turn the CR Reader ON.
c. Insert a cassette and see if one of the sensors is interfering
with the Screen Carrier movement.
Check/replace the Z0 J505 Sensor
according to
Sensors Tests” on Page 53
Yes No
Adjust the sensor position. Check that the Rotational motor
cable in the middle of the drum isn’t
interfering with the Linear Slide
movement.