“W0 Sensor at Wrong State”
Scenario
The CR Reader indicates that the W0 sensor (J511) is in failed state; it is in “1” position instead of “0”.
Hardware and Software Response
[1] Cassette is immediately ejected when inserted.
Fix
User Response
[1] Reset the CR Reader by turning it OFF and ON.
[2] Insert the cassette again, see if the error repeats.
Service Response
[1] Check/replace the W0 sensor (J511 on the Sensor board) (SK250013) according to
Sensors Tests” on
Page 53.
Yes No
Contact Service Continue using the CR Reader.