18 AT000115-02
“Homing cycle was not completed”
Scenario
The Linear motor moves too slowly or is stuck or does not reach the Right Limit sensor. (J515) (Time-out is 45
seconds)
Hardware and Software Response
[1] The error might include a mechanical noise.
[2] The CR Reader operating status is changed to “State stalled”.
[3] CR Reader will perform Homing loop until the problem is corrected.
Fix
User Response
[1] Reset the CR Reader by turning it OFF and ON.
[2] Does the error repeat?
Service Response
[1] Remove the CR Reader cover.
[2] Turn ON the CR Reader.
[3] Does the Linear Assembly move to the left and right when controlled from the Diagnostic screen?
Yes No
Contact Service Continue using the CR Reader.
Yes No
The Linear Assembly is sticking on the right side
and making loud noises. Check/replace J515
(“
Sensors Tests” on Page 53)
a. Check if the Linear Screw is stuck.
b. Verify that the coupling between the Linear Screw and the
Linear motor is firmly fastened from both sides.
c. Check/replace the Linear motor.