Software Error Messages
AT000115-02 29
Procedure Mapping Screens in Kodak QC User Interface
Error no. Error Message Text Possible Reason and Solution
1 This User does not have the required
permissions to access this utility.
User permission is denied.
2 Procedure Mapping Tool application is
already running.
If User tries to open Procedure Mapping tool more
than once.
3 Please fill a procedure name, User needs to write the Procedure Name
i n s i d e t h e t a b f r a m e .
4 A procedure with this name already exists Procedure name is duplicated.
5 Please select an item No item was selected when clicking on ADD button.
6 A restart is needed for all changes to take
effect
Restart the Kodak QC. Click [OK] then exit and
restart the Kodak QC.