Image Artifacts
AT000115-02 35
Section 6: Image Artifacts
White lines across image
White Lines Across Image Artifact Examples
Cause of Artifact
The Linear Slide gets stuck in the middle of a scanning procedure.
There might be a mechanical noise while scanning.
Note
When the Kodak Image Processing option is functioning this artifact is more severe.
Fix
Tighten the tension of the Slide Screw nut. This will also reduce the mechanical noise that accompanies scanning.
Required Tool:
12 mm wrench