36 AT000115-02
• 2.5 mm Allen tool
[1] Remove the three screws that fasten the Linear Screw bracket to the Optical Assembly.
Linear Screw Bracket
[2] Holding the Linear Screw bracket firmly, tighten the locknut clockwise by 2 mm
Tighten Nut
[3] Replace the three Linear Screw bracket screws.
[4] Perform a scanning procedure. If there is still a mechanical noise while scanning, or the artifact still appears,
repeat this procedure