Image Artifacts
AT000115-02 49
Vertical White Lines
Note
The lines are vertical to the scan. In the example on the right, the image was rotated.
Vertical White Lines Artifact Example
Cause
The USB Chipset on the computer Mother board is faulty.
Fix
[1] Check that the PC USB Chipset is the Intel model.
[2] Reinstall PC USB Chipset driver. It can be downloaded from the PC model website.
[3] Verify that the issue is with the PC by activating the CR Reader via another PC.
Note
If one of the steps solves the problem, there is no need to continue to the following steps.