50 AT000115-02
Image Partially Scanned
Partially Scanned Image Artifact Example
Cause
A W0 Sensor error message appears on screen during the scanning procedure. The Operator clicks [OK] to the
message and then the CR Reader immediately ejects the screen. Only part of the image has been scanned.
Fix
Workaround: Do not click [OK} before the CR Reader finishes scanning the screen.
Fix: Check the functionality of the W0 Sensor. (See
Sensors Tests” on Page 53) To fix the W0 Sensor go to “Plate
Didn’t Reach W0 Sensor” on Page 15 and “W0 Sensor at Wrong State”” on Page 16.