54 AT000115-02
Block the sensor signal
[5] If the sensor is not functioning properly, replace the sensor and repeat the test.
Note
To perform this test on J511 -W0 sensor: The sensor is divided into two parts; the Top is the Transmitter and the
Bottom is the Receiver. Insert a screen into the drum and check J511 sensor LED position change. During the screen
upload the LED should go from green to OFF.
Sensor blocked Sensor unblocked