Appendix 1: Tests
AT000115-02 55
Rollers Motor Test
[1] Remove the CR Reader cover.
[2] Turn the CR Reader ON.
[3] In Kodak QC software go to the Diagnostic screen.
[4] Click on <<->> to operate the Roller Motor control.
[5] Check visually that the rollers are spinning in the correct direction.
[6] Measure the voltage in the Roller Motor connector referring to the Roller Motor Sensor chart in the Point-of-
Care Service Manual.
Roller Motor Sensor Electrical connectors
[7] Check F2 on the Fuse Panel in the rear of the CR Reader.