7. Meter Settings [Setting] Screen
Operation
1.
Touch the item [Dark Calibration] button on page 2 of Setting.
"Perform Dark Calibration?" will be displayed.
7-4 Dark Calibration
Dark calibration is performed when new batteries are used, 24 hours passed since last
use or there is a big change in temperature between turning power OFF and ON again.
cept the cases above, dar calibration after power is sipped. ccordingly,
when there are sudden changes in temperature eperienced while using the meter
may effect measured values. When temperature changes are encountered, perform
dark calibration manually.
Light Selection Ring Status Indicator Chart
Icon Content
If either of icons appears, set
the Light Selection Ring
2
to
the dark calibration position.
This icon is the dark calibration
position.
Perform dark calibration after
making sure it is set to this
position.
Dark Calibration
Conrmation Screen
Setting Screen
Dark Calibration
Conrmation Screen