7. Meter Settings [Setting] Screen
2.
Turn the Light Selection Ring 2 to set to the dark calibration
position ( ).
The status bar will display the dark calibration Icon.
3.
Touch the [Yes] button.
"Dark calibration in progress. Please wait" and the status bar will appear while
meter's measuring system is being calibrated.
To return to Setting without performing dark calibration, touch the [No] button.
NOTICE
While the message "Dark calibration in progress. Please wait." or the
progress bar is displayed, do not turn the power off. Otherwise, the
meter may be damaged.
The dark calibration is completed.
Dark Calibration
Process Screen
Dark Calibration
Conrmation Screen
[No] Button[Yes] Button
Status Bar
Light Selection Ring