146
7. Meter Settings [Setting] Screen
When the Following Screen is Displayed
If the Light Selection Ring
2
is set to Range H
( ) or Range L ( ), dark calibration
cannot be performed.
Set the Light Selection Ring
2
to the dark calibration
position ( ), and perform dark calibration.
If Dark calibration was performed, but it could not be
completed normally.
Set the Light Selection Ring
2
to the dark calibration
position ( ) again, and perform dark
calibration again.
NOTE
●
Dark calibration can be done by turning the
Light Selection Ring
2
to set to the dark
calibration position ( ) in Measuring.
●
Dark calibration is performed when new
batteries are sed, 24 hours passed since last
use or there is a big change in temperature
between turning power OFF and ON again.
cept the cases above, dar calibration after
power ON is skipped.
Light Selection Ring
2
Dark Calibration Position
Conrmation Screen
Dark Calibration
Failure Screen