24
4. Setting the Measuring Conditions
4-1-2
Selecting the Exposure Time (Ambient Modes Only)
Set a posure Time for ambient measurement.
Operation
1.
T
ouch the [Exposure Time] indication on the Measurement screen.
Touch the [Auto], [0.1 sec] or [1.0 sec] button.
2.
Touch the [OK] button.
Conrms the settings, and returns to the previous Measurement screen.
Touch Cancel to return to the previous measurement screen without setting.
Exposure time is set.
NOTE
In uto mode, the CL-70F automatically selects among measuring times,
determined by the illumination available, to achieve a proper result in a convenient way.
Two ed reading times are available to enable eact comparison of multiple
measurements.
When measuring high illuminance levels, set the posure Time to 0. sec.
When taing measurements in low Illuminance, set the posure time for .0 sec.
NOTICE
.1 sec and 1. sec may not cover some measurement range of illumination
and [OVER] or [UNDER] appears. In this case, set Exposure time to "Auto".
Measuring Screen Exposure Time Screen
[Cancel]
Button
[OK]
Button
The selected
button will
be displayed
encircled in
blue.