390
10.2 Error codes and countermeasures
Error code Displayed message Problems/countermeasures
C37350
C37351
Internal error occurred in DICOM communication.
Contact a service representative.
Contact Konica Minolta technical representatives.
C37352
Failed to get DICOM communication setting. Please
conrm communication setting.
C37353
Failed in communication. Please conrm communica-
tion setting and server is working.
C37354
Failed in A-ASSOCIATE. Please conrm communica-
tion setting and server is working.
C37355
Rejected from server in A-ASSOCIATE due to invalid
setting. Please conrm communication setting.
C37356
Internal error occurred in DICOM communication.
Contact a service representative.
C37357 It failed in the communication of release.
C37358 A-ABORT was received. Check the network and destination connected to.
C3736*
C3737*
Internal error occurred in RIS-OUT communication.
Contact a service representative.
Data transfer failed or error occurred at the communi-
cation data creation.
(1) Please check the connection destination.
(2) Restart the device.
(3) Please check the setting again.
C37400 -
C37402
Conversion error
(1) Please check the connection status.
(2) Restart the device.
C37403 -
C37409
C3741*
C37420
C37421
Health check error
C37422 -
C37430
Scheduled check error
C37431 -
C37438
C3744*
Termination error
C37454 -
C37456
Study performing error
C37457 -
C37459
C3746*
C37470 -
C37472
Error in study completed notication to RIS
C37473 -
C37479
C3748*
C37495
Patient search error
C3760*
Internal error occurred in RIS-OUT communication.
Contact a service representative.
Data transfer failed or error occurred at the communi-
cation data creation.
(1) Please check the connection destination.
(2) Restart the device.
(3) Please check the setting again.
C3761*
Internal error occurred in RIS-IN communication.
Contact a service representative.
C37614
MWM C-FIND request aborted because it exceeded
the maximum number of search responses.
Contact Konica Minolta technical representatives.
C37662
Internal error occurred in RIS-OUT communication.
Contact a service representative.
Error occurred at the data transfer, the communication
data creation or the image creation.
(1) Output again.
(2) Please check the connection destination.
(3) Restart the device.
(4) Please check the setting again.