Setting the Range Manually
The measuring range is switched automatically from one to another
during measurement (5 ranges are available).
When you want to fix the range or you want to perform measurement of illuminance continuously
using the analog output function (see page 28), set the measuring range as follows.
Manual/auto measuring ranges
Range
Measuring range (unit: lx) Measuring range (unit: fcd)
Manual Auto Manual Auto
1 0.00 29.99 0.00 29.99 0.000 2.999 0.000 2.999
2 0.0 299.9 25.0 299.9 0.00 29.99 2.50 29.99
3
0 2999 250 2999 0.0 299.9 25.0 299.9
4 0
0
2999
0
250
0
2999
0
0 2999 250 2999
5 0
00
2999
00
250
00
2999
00
0
0
2999
0
250
0
2999
0
* The above measuring ranges are effective only when CCF correction is not used.
* It is possible to find the currently selected range by the position of he displayed decimal point, “0 or “00”.
Setting Method
1
Have the instrument ready for operation as described in
“Measuring the illuminance” (page 13) and push out the hold
button (RUN).
z
The range can be switched in RUN state only.
2
Insert the Mini plug (optional) into the analog output terminal.
“MANU. will appear in the display section.