Why does the signal relay switch none over despite level change in KSR KUEBLER Switch?
- RRichard RodgersAug 5, 2025
The signal relay might not switch over due to several reasons: * The P3 adjustment may be incorrect, placing it below the switching point. * The transducer could be faulty, exhibiting mechanical or electrical defects. * The glass tip might be positioned too close to the opposite surface. To resolve this, adjust according to manual 4.2. Perform a circuit board test. If the circuit board test is faultless, inspect the glass top for damage or dirt. Clean or replace if necessary. For transducer changes, refer to chapter 3.1.