3− 7
Section 3 Operation
HEMATOCRIT
15ml
HEMATOCRIT
15ml
15ml
HEMATOCRIT
The settable speed range can be changed over by pulling-out or inserting-in of the
15mL metal tubes from or into the tube holder.
And, the lighting of 15 mL lamp or HEMATOCRIT lamp on the control Panel changes accordingly.
㸦㸯㸧
Hematocrit capillary tube
Keepallfourmetaltubesinthetubeholder.
The〈HEMATOCRIT〉lampturnson.
Settable range 300-12,000rpm
㸦㸰㸧
15 mL metal tubes
Whenanyofthe15mLmetaltubesispulledoutofthefourholes
inthetubeholder,(Whenanyofthe15mLmetaltubesispulled
outofthefourholesinthetubeholder,thecentrifugejudgesthat
theyareused.)the〈15mL〉lampturnson.
Settable range 300-4,000rpm
How to use HT/15 Rotor
15mL metal tubes
Tube holder
NOTE
By pulling-out or inserting-in of the metal tubes into the tube holder, "Speed and
time" which have been preset with the metal tubes being inserted into the tube
holder and "Speed and time" which have been preset with the metal tubes not
being inserted into the tube holder will be changed over automatically in the setting
operation.
In the memory operation, only one of settings of hematocrit or 15mL metal tube can
be stored in one "MEMORY" key.