5−3 How to Use Peak Hold Function
10
ms or
100
ms responsetimecanbeselectedforpeakhold
measurement.Makeselectionaccordingtoyourapplicationneeds.
(
1
)
Withthetransformer jaws clampedontotheconductor undertest,
slide the peak hold switch from the OFF position to the desired
peakresponsetimeposition.
(
2
)
Thepeakholddisplayreads
1
/
2
ofthepeakcurrentvalue.
Therefore,anRMS reading willbeobtained when the current
undertesthasasinusoidalwaveform.
(
3
)
SlidethepeakholdswitchbacktotheOFFpositionforareset.
Note
1
KEW
2413
Fusesananaloguepeakholdcircuittoensurea
quickresponsetoinputcurrent.Becauseofthenatureofthis
circuit,thepeakholdreadingmaygraduallyfallor,inarare
case,rise withtime. Thisis likelyto beapparent whenthe
instrument is used in a high temperatureand high humidity
environment.Therefore,theinstrumentwillnotbesuitablefor
makingpeakmeasurementoveran extended periodoftime.
Incaseofsuchaneed,connectarecordertotheinstrument
viatheanalogueoutputterminal.