INSTALLATION: (Con’t)
INSTALLATION OF GAS FILTER INTERFACE ASSEMBLY KIT #018-514-00:
NOTE: Models 315, 315-1, 317, 317-1 require this assembly to be used for both CO2
and air.
KIT #018-514-00 INCLUDES:
4, Hose Clamps (#250-331-00)
1, In-Line Filter (#525-034-00)
1, 10 ft. (4.57m) length of 1/4” (6.35mm) ID PVC tubing (#720-418-00)
INSTALLATION PROCEDURE:
See next page for illustration.
Turn the unit power switch to the OFF position.
Connect tubing to tank gas outlet.
Connect 1/4” (6.35 mm) ID tubing to hose barb on one side of in-line filter
and secure with hose clamp; connect tubing to in-line filter hose barb on
its other side and secure with hose clamp.
314/316 series incubators, connect tubing with in-line filter in place to the
appropriate gas inlet fitting on the back of the unit. 315/317 series
incubators require this assembly to be connected to both air and gas inlet
fittings.
Set the supply regulators at 4 psi.
Check all of the connections for leaks.
Turn the unit power switch to the ON position.
WARNING: HIGH CONCENTRATIONS OF CARBON DIOXIDE PRODUCE
METABOLIC ABNORMALITIES, DISTURBANCES OF THE CENTRAL
NERVOUS SYSTEM AND CARDIAC INSTABILITY—UNCONSCIOUSNESS
MAY OCCUR AT CONCENTRATIONS ABOVE 10%.