12
If you are not sure that the found heart rate is the fetal, you can instruct the mother
to hold the pads to measure mother’s HR. If the two numbers are similar, then the HR
found by the ultrasound sensor might be the maternal HR.
87
87
87
85
135
12.12.2014
CONFIGURATION
Lorem Ipsum
Ox2O
195
?
?
? ?
95
15:36
134
83
97
97
135
134 134
Use sequence
15:36
134
Instruct the mother to hold the pads.
Compare the HR found by the ultrasound sensor with the maternal HR.
PRACTICE
12 13
Refer to page 32 for details about the maternal HR display.
The ultrasound sensor might measure or be affected by mother's HR, a twin (multiples)
or excessive patient movement/handling. In case the device displays unreliable or erratic
FHR, try to reposition the ultrasound transducer or use alternative monitoring devices/
methods.